DEPUY MITEK LLC US BIO-INTRAFIX TIBIAL SHEATH - SMALL; SOFT-TISSUE ANCHOR, BIOABSORBABLE Back to Search Results
Model Number 254629
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/22/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Additional narrative: udi: (b)(4).To date, the device has not been returned.If the device or further details are received at a later date, a supplemental medwatch will be sent.
 
Event or Problem Description
It was reported by a healthcare professional in china that during an anterior cruciate ligament reconstruction surgery on (b)(6) 2022, it was observed that the bio-intrafix tibial sheath small device was broken off.Removed all the broken parts.Another device was used to complete the surgery.There were no adverse consequences to the patient.No additional information could be provided.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.D9, h3, h6: the actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary: the complaint device was received and evaluated.Upon visual inspection, the anchor was not returned, however the sheath which is part of the kit was returned.The sheath is broken in two separate pieces.A manufacturing record evaluation was performed for the finished device 8l56966 number, and no non-conformances were identified.As part of depuy synthes mitek quality process all devices are manufactured, inspected, and released to approved specifications.According with the inspection, this complaint cannot be confirmed.The customer reported a broken anchor, however they only returned the sheath.A possible root cause for the broken sheath can be attributed a failure of using the guidewire while insertion.As per ifu; failure to insert the screw along the tunnel axis may result in device damage.Guidewire use is recommended.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to monitor additional complaint information for potential safety signals through complaint trending as part of post market surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIO-INTRAFIX TIBIAL SHEATH - SMALL
Common Device Name
SOFT-TISSUE ANCHOR, BIOABSORBABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key15846219
Report Number1221934-2022-03701
Device Sequence Number12191860
Product Code HWC
UDI-Device Identifier10886705013258
UDI-Public10886705013258
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K032167
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2024
Device Model Number254629
Device Catalogue Number254629
Device Lot Number8L56966
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/08/2022
Supplement Date Received by Manufacturer12/05/2022
01/27/2023
Initial Report FDA Received Date11/22/2022
Supplement Report FDA Received Date12/06/2022
01/30/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/16/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-