| Model Number |
COBAS 8000 ISE |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/28/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
There was an allegation of questionable gen.2 ise indirect for k results for 1 patient sample on a cobas 8000 cobas ise module.On (b)(6) 2022, two samples were collected from the patient at the same time.On (b)(6) 2022, the initial k result from the first sample was 5.9 mmol/l.The sample was repeated two more times and the result both times was 6.0 mmol/l.The result of 6.0 mmol/l was reported outside of the laboratory.On (b)(6) 2022, the second sample was repeated and the k result was 5.2 mmol/l.The repeat result of 5.2 mmol/l was deemed correct.The k electrode lot number is j8377.The expiration date was requested but not provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer stated that qc recovery was acceptable.The investigation determined that the cause of the event is consistent with pre-analytic sample handling issues.The investigation did not identify a product problem.The root cause of the event could not be determined.
|
| |
|