| Model Number |
884838070974 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Vascular Dissection (3160); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/09/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
This case was reviewed and investigated according to the manufacturer¿s policy.Date of birth not available.Not applicable for this device.Blocks d6 & d7: not applicable for this device.Blocks e1 & g2: country is germany.Block g2: study name - learn registry.Blocks h3 & h6: the altatrack guidewire was discarded by the facility, thus no returned product investigation was performed.Blocks h7 & h9: do not apply to this submission.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2021, an altatrack guidewire was used in a bevar procedure in the abdominal aorta plus side branches.The altatrack guidewire was used along with non-philips conventional guidewires in combination with conventional catheters.No patient injury or device malfunction was reported at the time of the procedure.However, 8 days post procedure on (b)(6) 2021, a dissection was noted in the descending aorta (celiac trunk, stenotic >50%, not calcified <50%), but no intervention was performed.This adverse event is being submitted in abundance of caution because it could not be determined if the altatrack guidewire caused or contributed to the dissection.
|
| |
|
Additional Manufacturer Narrative
|
|
Blocks b5/g3: additional information was provided by the site on 23nov2022.Block d10: terumo guidewire was listed as a concomitant device in the initial mdr.Block f10: the hecc and heic complaint codes were updated to reflect no patient injury: hecc code 3160 (vascular dissection) was replaced with code 4582 (no clinical signs, symptoms or condition).Heic code 4164-serious injury/ illness/ impairment) was replaced with code 2199 (no health consequences or impact).Block h2: correction- this event is no longer reportable based on the additional information received.Block h6: added conclusion code 4310 (caused traced to non-device related factors).Code 67 (no problem detected) from the initial mdr remains applicable.
|
| |
|
Event or Problem Description
|
|
This adverse event is no longer reportable for a dissection.Additional information confirmed the altatrack device did not cause the dissection.The task was completed with a terumo guidewire.
|
| |
|