| Model Number |
14679-05 |
| Medical Device Problem Code |
Firing Problem (4011)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/01/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that an arteriotomy closure of the left common femoral artery was attempted using a starclose device after a diagnostic left leg angiogram procedure with a 6f sheath.Reportedly, the device didn't deploy correctly as the clip was deployed into the exchange sheath [firing problem].Manual arterial compression was used to achieve hemostasis.There was no adverse patient sequela and no reported clinically significant delay in the procedure or therapy.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history identified no similar incidents from this lot.The investigation was unable to determine a conclusive cause for the reported difficulties; however, the treatment appears to be related to circumstances of the procedure.Based on the information reviewed, there is no indication of a product quality issue with respect to manufacture, design or labeling.
|
| |
|