It was reported that on (b)(6) 2022, a small flexnav delivery system was selected to implant a device.During the procedure, as the physician advanced the delivery system, some resistance was observed and the delivery system was immediately removed.A kink was observed in the end of the valve capsule.The physician alleged that the kink was caused by an incorrectly placed proglide on the guidewire which caught on the device whilst he inserted it in the delivery system.Prior to the procedure, there was no damage done to the product box or the device itself and the device no damaged due to handling during preparation was reported.The device was exchanged with a new flexnav delivery system that deployed the valve successfully with no further issues.There was prolonged procedure time and the patient needed an extra dose of contrast and radiation.The patient also had some vascular complications at the end of the case, no treatment was provided.The physician allege that a small femoral anatomy along with the proglide caused the issue.The patient remain hemodynamically stable throughout the procedure,.
|