• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MMJ SA DE CV (USD) SHILEY; TUBE TRACHEOSTOMY AND TUBE CUFF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MMJ SA DE CV (USD) SHILEY; TUBE TRACHEOSTOMY AND TUBE CUFF Back to Search Results
Model Number 6.0PELF
Medical Device Problem Code Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/01/2022
Type of Reportable Event Malfunction
Event or Problem Description
According to the reporter, there was difficulty on inserting and removing the device introducer.There was no patient involvement.
 
Additional Manufacturer Narrative
Concomitant medical products: 6.0pelf (6.0pelf 6.0mm id paed lng lgecufflessx1.Lot number: 22c1306jzx); 6.0pelf (6.0pelf 6.0mm id paed lng lgecufflessx1.Lot number: 22c1306jzx); 6.0pelf (6.0pelf 6.0mm id paed lng lgecufflessx1.Lot number: 22c1306jzx); 6.0pelf (6.0pelf 6.0mm id paed lng lgecufflessx1.Lot number: 22c1306jzx); 6.0pelf (6.0pelf 6.0mm id paed lng lgecufflessx1.Lot number: 22c1306jzx); 6.0pelf (6.0pelf 6.0mm id paed lng lgecufflessx1.Lot number: 22c1306jzx); 6.0pelf (6.0pelf 6.0mm id paed lng lgecufflessx1.Lot number: 22c1306jzx); 6.0pelf (6.0pelf 6.0mm id paed lng lgecufflessx1.Lot number: 22c1306jzx); 6.0pelf (6.0pelf 6.0mm id paed lng lgecufflessx1.Lot number: 22c1306jzx); 6.0pelf (6.0pelf 6.0mm id paed lng lgecufflessx1.Lot number: 22c1306jzx).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that there was difficulty on inserting and removing the device obturator.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.The instructions included with this device provide the following guidance: prior to tube insertion, check that the obturator can be inserted and removed without difficulty.A thin film of water soluble lubricant should be applied to the tube and protruding portion of the obturator to facilitate ease of insertion.Insert the tube with the obturator fully seated in place.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHILEY
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
MMJ SA DE CV (USD)
ave henequen no 1181 desarroll
ciudad juarez 32590
MX  32590
Manufacturer (Section G)
MMJ SA DE CV (USD)
ave henequen no 1181 desarroll
ciudad juarez 32590
MX   32590
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key15846862
Report Number2936999-2022-01063
Device Sequence Number5027268
Product Code JOH
UDI-Device Identifier20884521779898
UDI-Public20884521779898
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K182861
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number6.0PELF
Device Catalogue Number6.0PELF
Device Lot Number22C1306JZX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/22/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/04/2022
Supplement Date Received by Manufacturer02/20/2023
Initial Report FDA Received Date11/22/2022
Supplement Report FDA Received Date02/28/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/18/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10
Patient SexUnknown
-
-