| Model Number |
TRF-450 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
10/24/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Healthcare professional reported right side rupture.Device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.A review of the device history record has been completed.No deviations or non-conformance's noted.Reason for reoperation: rupture.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the device analysis grid, the assessments of the complaint are: rupture: observed, opening on the posterior side assessed as surgical impact opening.(shell thickness was within specification) additional observations: no penetrating nick observed on the device.Crease fold observed on the device.No further actions are required as the device was implanted.
|
| |
|
Event or Problem Description
|
|
Device has been explanted.
|
| |
|