| Model Number |
G5A-01A |
| Medical Device Problem Codes |
Unable to Obtain Readings (1516); Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during biomed testing, the device displayed an unknown critical error message.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental medwatch report is reporting the evaluation of the device and correcting information submitted on the initial medwatch report.Please reference section b5 and h6 (device problem code).Evaluation: the device was returned to zoll medical corporation.The customer's report of "the device was unable to detect the attached electrode pads" was not replicated or confirmed.The device was put through extensive testing without duplicating the report.Review of the device log found no evidence of the report.The customer's report of "the device prompted an "hv dump failure" message" was verified and attributed to the main board.The main board was replaced to remedy the report.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during biomed testing, the device was unable to detect the attached electrode pads and prompted an "hv dump failure" message.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|