• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) JOT DX¿ ICM; IMPLANTABLE CARDIAC MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) JOT DX¿ ICM; IMPLANTABLE CARDIAC MONITOR Back to Search Results
Model Number DM4500
Device Problems Signal Artifact/Noise (1036); Migration or Expulsion of Device (1395)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/03/2022
Event Type  Injury  
Event Description
It was reported that after the incision was closed and the drape removed, the device was found to be lying on the patient's outer chest and not in the pocket.Noise was noted on the iegm due to the device no longer being in the patient.The device was explanted and replaced.The patient was stable.
 
Manufacturer Narrative
Sensing problem and device migration issue were not confirmed.The device was received in normal working conditions with battery voltage above eri.Analysis performed, including sensing test at field and high sensitivity settings indicated normal device functionality.Longevity assessment was performed, and device was in the normal range of operation with appropriate remaining longevity.
 
Manufacturer Narrative
Correction g3: the date for the previous report follow-up 1 should have been dec 20, 2022.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JOT DX¿ ICM
Type of Device
IMPLANTABLE CARDIAC MONITOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key15847306
MDR Text Key304170569
Report Number2017865-2022-46301
Device Sequence Number1
Product Code MXC
UDI-Device Identifier05415067040046
UDI-Public05415067040046
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K163407
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 01/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/08/2023
Device Model NumberDM4500
Device Catalogue NumberDM4500
Device Lot NumberP000148995
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/09/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/03/2022
Initial Date FDA Received11/22/2022
Supplement Dates Manufacturer Received12/15/2022
01/09/2023
Supplement Dates FDA Received01/05/2023
01/10/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/11/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age20 YR
Patient SexFemale
-
-