| Model Number |
3224 |
| Medical Device Problem Code |
High impedance (1291)
|
| Health Effect - Clinical Code |
Inadequate Pain Relief (2388)
|
| Date of Event |
05/19/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Date of event is estimated.During processing of this complaint, attempts were made to obtain patient weight.The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
| |
|
Event or Problem Description
|
|
It was reported the patient experienced ineffective therapy.Lead diagnostics indicated high impedances.As a result, surgical intervention occurred wherein the lead was explanted and replaced to address the issue.
|
| |
|