AVANOS MEDICAL INC. CORFLO NASOGASTRIC/NASOINTESTINAL FEEDING TUBE WITH STYLET WITH ENFIT; DH CPK NG TUBES Back to Search Results
Model Number 42-9432
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 10/15/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product involved in the report has been returned and is being processed for evaluation.A review of the device history record is in-progress.All information reasonably known as of 22-nov-2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint: (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
 
Event or Problem Description
It was reported " an alarm rang and it was found that ng tube was clogged.A nurse tried to flush the ng tube with lukewarm water using 10cc syringe, but the water did not run.After flushing the tube a few times, the tube was torn under y-connector.Another tear was found around at 20cm.Also the area was stretched.".
 
Additional Manufacturer Narrative
All information reasonably known as of 29-nov-2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
 
Event or Problem Description
Additional information received 28-nov-2022 stating "it was not sure how many times the user had administered nutrition" by the time the incident occurred.There were no reports of injury nor medical intervention.
 
Additional Manufacturer Narrative
The device history record for the reported lot number, 30148032, in this complaint was reviewed and the material was produced according to the manufacturing procedures and met the quality requirements.The sample was received without the original packaging.No stylet wire received.The incident reported was confirmed per sample evaluation, there was damage/ splitting of tubing below the y-connector, the entirety of tubing was unable to flush.When trying to flush the device from the split opening on the tube with syringe, resistance was felt while flushing.Materials appeared stuck inside the tube and was unable to be flushed out of the tube after multiple attempts.There was also another split/tear identified approximately at the 20cm marking with a ballooned shape of the tube.A review of the instructions for use (ifu) indicated the following: tube maintenance: follow your institution/facility/hospital protocol or clinician¿s order.It is recommended the tube be irrigated every 4 hours with up to 20 ml of water (up to 10 ml for infants or children) before and after medication administration or when feeding formula is interrupted.Warning: vigorous syringe force should not be used to irrigate, administer liquids or unblock the tube.The feeding tube should be monitored, regularly assessed, and replaced when clinically indicated based on functionality and patient condition.Root cause could not be determined.All information reasonably known as of 27-nov-2023 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CORFLO NASOGASTRIC/NASOINTESTINAL FEEDING TUBE WITH STYLET WITH ENFIT
Common Device Name
DH CPK NG TUBES
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
Manufacturer (Section G)
AVENT SA DE R.L. DE C.V. (AVENT 1)
circuito industial no.40
colonia obrera
nogales, cp
Manufacturer Contact
lisa clark
5405 windward parkway
alpharetta, GA 30004
4704485444
MDR Report Key15848035
Report Number9611594-2022-00144
Device Sequence Number20464277
Product Code KNT
UDI-Device Identifier10680651471687
UDI-Public10680651471687
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K821906
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/08/2026
Device Model Number42-9432
Device Catalogue NumberN/A
Device Lot Number30148032
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/17/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/31/2022
Supplement Date Received by Manufacturer11/28/2022
01/04/2023
Initial Report FDA Received Date11/22/2022
Supplement Report FDA Received Date11/29/2022
01/30/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/12/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1