Model Number 190858
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/18/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity.
 
Event or Problem Description
A user facility biomedical technician (biomed) reported an e.23 alarm that occurred at the beginning of a patient¿s hemodialysis (hd) treatment on a fresenius 2008t machine.The staff was unable to clear the alarm and the patient¿s blood could not be returned.The patient¿s estimated blood loss (ebl) was 300 ml.It was confirmed there was no patient injury, no adverse effects were experienced, and no medical intervention was required as a result of the reported event.The patient was moved to a different machine where they completed treatment after being re-setup with new supplies.To resolve the reported machine issue, the blood pump motor was replaced.The biomed stated there were no issues with the disposable supplies that were in use at the time of the event.A sample is not expected to be returned for manufacturer evaluation.
 
Additional Manufacturer Narrative
Plant investigation: no parts were returned to the manufacturer for physical evaluation and an on-site evaluation was not performed by a fresenius field service technician (fst).However, the manufacturer was able to determine a causal relationship between the objective evidence provided by the customer and the reported event.A records review was performed on the reported serial number.An investigation of the device manufacturing records was conducted by the manufacturer.There were no non-conformances, or any associated rework identified during the manufacturing process which could be related to the reported event.In addition, the device history record (dhr) review confirmed the results of the in-progress and final quality control (qc) testing met all requirements.Based on the provided information the reported event has been confirmed.
 
Event or Problem Description
A user facility biomedical technician (biomed) reported an e.23 alarm that occurred at the beginning of a patient¿s hemodialysis (hd) treatment on a fresenius 2008t machine.The staff was unable to clear the alarm and the patient¿s blood could not be returned.The patient¿s estimated blood loss (ebl) was 300 ml.It was confirmed there was no patient injury, no adverse effects were experienced, and no medical intervention was required as a result of the reported event.The patient was moved to a different machine where they completed treatment after being re-setup with new supplies.To resolve the reported machine issue, the blood pump motor was replaced.The biomed stated there were no issues with the disposable supplies that were in use at the time of the event.A sample is not expected to be returned for manufacturer evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2008T HEMODIALYSIS SYSTEM WITHOUT CDX
Common Device Name
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
Manufacturer (Section G)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
Manufacturer Contact
jessica trujillo
920 winter st
waltham, MA 02451
6174175172
MDR Report Key15848351
Report Number0002937457-2022-02034
Device Sequence Number5116246
Product Code KDI
UDI-Device Identifier00840861100903
UDI-Public00840861100903
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K150708
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 11/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number190858
Device Catalogue Number190858
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Device AgeMO
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/01/2022
Supplement Date Received by Manufacturer11/23/2022
Initial Report FDA Received Date11/22/2022
Supplement Report FDA Received Date11/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/23/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
FRESENIUS BLOODLINES; FRESENIUS BLOODLINES; FRESENIUS OPTIFLUX DIALYZER; FRESENIUS OPTIFLUX DIALYZER
Patient Age47 YR
Patient SexMale
-
-