• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL SJM RIGID SADDLE RING; RING, ANNULOPLASTY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL SJM RIGID SADDLE RING; RING, ANNULOPLASTY Back to Search Results
Model Number RIGID SADDLE RING
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Pleural Effusion (2010); Heart Block (4444)
Date of Event 10/19/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
(b)(4).It was reported that on (b)(6) 2022, a 32mm rigid saddle ring was successfully implanted in a patient.On (b)(6) 2022, the patient had a third-degree atrioventricular (av) block and a pacemaker was implanted.On 05 november 2022, the patient presented to the emergency room with pleural effusion and surgical intervention was performed.The patient was reported to be in stable condition.
 
Additional Manufacturer Narrative
An event of third-degree atrioventricular (av) block and pleural effusion was reported.A returned device assessment could not be performed as the device remains implanted and was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.Based on the information received, the cause of the reported incident could not be conclusively determined., but may be caused by the non-abboot device which was used during the same procedure.There is no indication of a product quality issue with respect to the labeling design or manufacturing of the device.
 
Event or Problem Description
(b)(4).(b)(6).It was reported that on (b)(6) 2022, a 32mm rigid saddle ring was successfully implanted in a patient.On (b)(6) the patient had a third-degree atrioventricular (av) block and a pacemaker was implanted.On (b)(6) 2022, the patient presented to the emergency room with pleural effusion and surgical intervention was performed.It was also reported that a non-abbott device was implanted at time of 32mm rigid saddle ring implant.The patient received a tricuspid valve repair in the same procedure.The device used was from another manufacturer.The patient was reported to be in stable condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SJM RIGID SADDLE RING
Common Device Name
RING, ANNULOPLASTY
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL PUERTO RICO, INC. REG#2648612
20 b st caguas west park
caguas 00725
*   00725
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key15848492
Report Number2135147-2022-02189
Device Sequence Number17749705
Product Code KRH
UDI-Device Identifier05414734009454
UDI-Public05414734009454
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K042734
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Study,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberRIGID SADDLE RING
Device Catalogue NumberRSAR-32
Device Lot Number8480597
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 11/03/2022
Supplement Date Received by Manufacturer02/17/2023
Initial Report FDA Received Date11/22/2022
Supplement Report FDA Received Date02/20/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/18/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age71 YR
Patient SexMale
Patient Weight83 KG
Patient RaceWhite
-
-