• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL AMPLATZER TALISMAN PFO OCCLUDER; TRANSCATHETER SEPTAL OCCLUDER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL AMPLATZER TALISMAN PFO OCCLUDER; TRANSCATHETER SEPTAL OCCLUDER Back to Search Results
Model Number 9-PFO-3525
Medical Device Problem Code Difficult to Fold, Unfold or Collapse (1254)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/25/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An event of device deformity was reported.A returned device assessment could not be performed as the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.There were two complaints associated with any other devices including the current device on material deformation from the lot 8213789.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.
 
Event or Problem Description
It was reported that on (b)(6) 2022, a 35-25mm amplatzer talisman pfo occluder (8213789) was chosen for implant using a 9f amplatzer talisman delivery system (8157401).The device was tested prior to procedure and deployed normally, as expected.When attempting to deploy the pfo occluder, the right atrial disk deployed in a bulbous formation.A kink was noticed in the 9f delivery system (8157401) during the first delivery of the device.The device was not released from the delivery cable while in the patient.The device was retrieved and removed from the patient.There was no noted interaction with cardiac structures.An attempt was made to restore the normal shape while testing outside of the patient, but the device did not deploy in the correct formation.Therefore, it was replaced it with a new 35-25mm amplatzer talisman pfo occluder (8221783) and it deployed as expected.The replacement occluder was successfully implanted using a new 9f amplatzer talisman delivery system (8277458).There were no adverse patient effects.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER TALISMAN PFO OCCLUDER
Common Device Name
TRANSCATHETER SEPTAL OCCLUDER
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT MEDICAL REG# 2135147
5050 nathan ln n
plymouth MN 55442
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key15848496
Report Number2135147-2022-02175
Device Sequence Number6429586
Product Code MLV
UDI-Device Identifier05415067033345
UDI-Public05415067033345
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P120021
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2024
Device Model Number9-PFO-3525
Device Catalogue Number9-PFO-3525
Device Lot Number8213789
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/26/2022
Initial Report FDA Received Date11/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/09/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AMPLATZER TALISMAN DS, 9-TDS-09F45-80, 8157401
Patient SexMale
-
-