• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE FORMULA 180 SHAVER INTERNATIONAL KIT.; BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE FORMULA 180 SHAVER INTERNATIONAL KIT.; BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL Back to Search Results
Model Number 0375708500
Medical Device Problem Code Continuous Firing (1123)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/28/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the device was continuously activating.
 
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Additional Manufacturer Narrative
The device was received at the local service facility for investigation.The technical service report is attached(see communication log), and indicates: service code: a the repair includes: defective parts replacement, leakage current function test calvier rplcmt joint, membrane, revision.Probable root cause: breaking resistors on the flex circuitry due to stress on flex cable (due to manufacturing and assembly error combined with normal use stress).Broken handpiece cable, or conductors inside.No power from the console.Broken traces on the membrane switch.Membrane switch is corroded due to moisture ingress.Flattened buttons/ switches.Poor soldering or other electrical connection.Broken pin.Pcba component failure.Button press reaction force so high the buttons can't be pressed to deliver.Information.Console not delivering power to shaver (refer to rsk10642, rsk10686).Expired disposable after r.1.1.1.Use error.The reported failure mode will be monitored for future reoccurrence.
 
Event or Problem Description
It was reported that the device was continuously activating.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FORMULA 180 SHAVER INTERNATIONAL KIT.
Common Device Name
BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
lucas wolski
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key15848536
Report Number0002936485-2022-00698
Device Sequence Number12110121
Product Code GFA
UDI-Device Identifier07613327057645
UDI-Public07613327057645
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0375708500
Device Catalogue Number0375708500E
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/28/2022
Supplement Date Received by Manufacturer10/28/2022
Initial Report FDA Received Date11/22/2022
Supplement Report FDA Received Date01/21/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/01/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-