MENTOR TEXAS MENTOR CPX 4 BREAST TISSUE EXPANDER WITH SUTURE TABS; EXPANDER, SKIN, INFLATABLE
|
Back to Search Results |
|
| Model Number |
3549211 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/04/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a mentor cpx 4 breast tissue expander with suture tabs 275cc device deflated during an unspecified breast surgery procedure.The surgeon found a pin hole on the device and did not use it.The surgery was completed with a different unspecified tissue expander.There was no reported patient consequence.
|
| |
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: n/a.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number:(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On 02-dec-2022, the mentor failure analysis lab received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On 13-mar-2023, mentor completed the investigation on the suspect medical device.Device evaluation summary: according to the information received, during the procedure, the customer reported a defect (pinhole) in the tissue expander.The product was returned to mentor for evaluation.Mentor conducted a visual inspection and leak testing of the returned device.Visual analysis of the returned device determined that an excess material was found inside the anterior view of the device, measuring approximately 5.0 mm² (.05 cm²).The tissue expander was received without packaging.Leak testing was performed, in accordance with mentor procedures, and no leak sites were detected during the analysis.Based on the information currently available, a product issue was identified during the investigation of the sample received.This product issue will be addressed through mentor¿s quality system.While this complaint was initiated for a defect (pinhole) on the tissue expander, through the assessment, it was identified to be an excess of silicone material as part of the manufacturing process that will not cause injury and is unlikely to render the product unusable or difficult to use.There are inspections at various stages of the manufacturing process.However, excess silicone is a normal variation that can occur.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.H6 investigation findings ¿ c22: cosmetic defect.Manufacturer¿s reference number: (b)(4).
|
| |
|