Model Number PM3242
Device Problem Failure to Disconnect (2541)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/07/2022
Event Type  malfunction  
Event Description
Related manufacturer report number: 2017865-2022-46385.During a routine device replacement procedure, it was noted the right ventricular (rv) lead was unable to be disconnected from the device header.The rv lead was capped and replaced to resolve the event.The patient was stable.
 
Manufacturer Narrative
The reported event of the difficulty removing the ventricular lead from the header was confirmed.Analysis revealed this device has the original scalable brady pacemaker (sbp) header design, which is known to have lead insertion and removal difficulties.The original sbp header required extra above normal forces to insert and to remove leads from the connector ports.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLURE QUADRA RF CRT-P
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key15849310
MDR Text Key306829481
Report Number2017865-2022-46384
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734507912
UDI-Public05414734507912
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P030035
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/22/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2016
Device Model NumberPM3242
Device Catalogue NumberPM3242
Device Lot Number4819722
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/14/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/03/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
TENDRIL ST
-
-