MAUDE Adverse Event Report: IMPLANT DIRECT SYBRON MANUFACTURING LLC LEGACY3 IMPLANT; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

IMPLANT DIRECT SYBRON MANUFACTURING LLC LEGACY3 IMPLANT; DENTAL IMPLANT Back to Search Results
Catalog Number 864710
Medical Device Problem Code Osseointegration Problem (3003)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 11/07/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Per complaint (b)(4), after clinical procedure, patient experienced failure of implant to osseointegrate.
 
Additional Manufacturer Narrative
Other relevant history, including preexisting medical conditions were not provided.If the requested information becomes available, a supplementary report will be submitted.Patient identifier, age, oral hygiene and bone quality are unknown implant and explant dates are unknown.The implant failure modes, failure/ loss of osseo-integration and lack of primary stability are not product related and rather are attributed to patient contraindications, conditions, or clinician error in surgical protocol.No product investigation or corrective actions required.Complaints will continue to be trended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEGACY3 IMPLANT
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
IMPLANT DIRECT SYBRON MANUFACTURING LLC
3050 east hillcrest drive
thousand oaks CA 91362
Manufacturer Contact
marcos gutierrez
3050 east hillcrest drive
thousand oaks, CA 91362
8184443315
MDR Report Key15850150
Report Number3001617766-2022-05742
Device Sequence Number5116837
Product Code DZE
UDI-Device Identifier10841307102571
UDI-Public10841307102571
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
K090234
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Dentist
Remedial Action Inspection
Type of Report Initial
Report Date (Section B) 11/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/12/2021
Device Catalogue Number864710
Device Lot Number77997
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/08/2022
Initial Report FDA Received Date11/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/13/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
Patient Weight64 KG
-
-