Additional Manufacturer Narrative
Ethnicity: unknown/asked but unavailable.Race: unknown/asked but unavailable.Date of event: the exact date is unknown.The best estimate is between the implant and explant dates, jun 22, 2022 through oct 26, 2022.Device evaluation: the device was not returned at the manufacturing site; therefore, product testing could not be performed and the customer¿s reported complaint could not be verified.Manufacturing record review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured according to specifications.A search of complaints related to this production order (po) was performed.The search of complaints revealed that no similar complaint for this production order number has been received.Conclusion: as a result of the investigation there is no indication of a product quality deficiency, and the reported issue could not be confirmed.Attempts were made to obtain missing information; however, the account did not have the information.All pertinent information available to johnson & johnson surgical vision, inc., has been submitted.
 
Event or Problem Description
It was reported that the intraocular lens (iol) was explanted from the patient¿s right eye due to them experiencing halos, glare and mechanical issues.There was no capsule tear, unplanned vitrectomy, incision enlargement or sutures required.Another johnson & johnson lens, model dcb00 23.5 diopter was implanted as a replacement.Reportedly, the patient post-op is happy at 20/20 vision.No further information was provided.
 
Additional Manufacturer Narrative
Section d9: device available for evaluation? yes.Section d9: date returned to manufacturer: may 11, 2023.Section h3: evaluated by manufacturer: yes.Device evaluation: the suspect iol was received cut in half.The lens was cleaned and visual inspection under magnification found no issues that could contribute to the event reported.No further evaluation was performed.Conclusion: as a result of the investigation there is no indication of a product quality deficiency, and the reported issue could not be verified.Please note that the information reported in sections d2 (common device name), h.6 (health effect -clinical code and medical device problem code) and section g.1 (manufacturer contact e-mail) is information that remains unchanged from the initial emdr report.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
LENS, MULTIFOCAL INTRAOCULAR
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key15850283
Report Number3012236936-2022-02728
Device Sequence Number6449888
Product Code MFK
UDI-Device Identifier05050474560758
UDI-Public(01)05050474560758(17)251222
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/22/2025
Device Model NumberZLB00
Device Catalogue NumberZLB00U0240
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/31/2022
Supplement Date Received by Manufacturer05/19/2023
Initial Report FDA Received Date11/23/2022
Supplement Report FDA Received Date06/07/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/22/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
Patient Weight93 KG
-
-