MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. DORADO; PTA BALLOON DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. DORADO; PTA BALLOON DILATATION CATHETER Back to Search Results
Model Number DR135104
Medical Device Problem Codes Leak/Splash (1354); Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/01/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an angioplasty procedure via left brachial artery, the pta balloon allegedly leaked and a pinhole rupture was noted.The procedure was completed using another device.There was no reported patient injury.
 
Additional Manufacturer Narrative
Manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: the physical device was not returned for evaluation.One photo was provided.The photo shows a health professional holding the dorado balloon which is in an inflated condition.No evidence of rupture can be seen and no anomalies noted.Therefore, the investigation is inconclusive for the reported pinhole balloon rupture as no evidence of the rupture could be noted.A definitive root cause for the reported pinhole balloon rupture could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.Expiry date: 05/2024.
 
Additional Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: the physical device was not returned for evaluation.Three photos were provided and reviewed.The first photo showed a health professional holding the dorado balloon which was in an inflated condition.The second photo showed a drop of liquid falling from the balloon.The third photo showed a droplet at the edge of the balloon.However, the exact source of the leak could not be determined.No evidence of rupture could be seen and no other anomalies were noted.Therefore, the investigation is inconclusive for the reported pinhole balloon rupture and confirmed for the identified leak as droplets were seen falling from the balloon from the photo review but the exact source of the leak could not be determined.A definitive root cause for the reported pinhole balloon rupture and identified leak could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 05/2024).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
 
Event or Problem Description
It was reported that during an angioplasty procedure via left brachial artery, the pta balloon allegedly leaked and a pinhole rupture was noted.The procedure was completed using another device.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DORADO
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
FUTUREMATRIX INTERVENTIONAL
1605 enterprise street
athens 75751
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key15850769
Report Number2020394-2022-00915
Device Sequence Number6449901
Product Code LIT
UDI-Device Identifier00801741059568
UDI-Public(01)00801741059568
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K072283
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberDR135104
Device Catalogue NumberDR135104
Device Lot Number93SF0089
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/27/2022
Supplement Date Received by Manufacturer04/12/2023
Initial Report FDA Received Date11/23/2022
Supplement Report FDA Received Date04/24/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age62 YR
Patient SexFemale
Patient Weight66 KG
-
-