Model Number CCWET0
Medical Device Problem Code Scratched Material (3020)
Health Effect - Clinical Code Blurred Vision (2137)
Date of Event 10/25/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The product was returned for analysis.Intraocular lens (iol) returned in a plastic container in balanced salt solution (bss) solution.Iol is air dried.Light scratches are observed on iol optic.Toric marks confirmed.Carton on primary label also returned showing.No secondary label set returned.The lens was sterilized and measured by r&d engineer (b)(6) in r&d pilot line and was confirmed as company lens.Photo shows implanted iol.3 dots are visible on the iol optic at the gusset location.Toric marks confirmed.The root cause is deemed to be manufacturing related.Incorrectly labelled product was released.Based on the results from the product history record, the products met release criteria.There have been no other complaints for this lot.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A physician reported that during an intraocular lens (iol) implant procedure, surgeon reported a non-toric lens had toric markings.Noted upon insertion of lens and surgeon was unsure of what to do.Left lens implanted.On one day post op patient is seeing 20/200 and unable to see intermediate at all.Additional information was requested and received the iol was exchanged in a secondary procedure following the initial procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON VIVITY EXTENDED VISION UV ABSORBING IOL
Common Device Name
EXTENDED DEPTH OF FOCUS INTRAOCULAR LENS
Manufacturer (Section D)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI  00000
Manufacturer (Section G)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI   00000
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15851209
Report Number9612169-2022-00637
Device Sequence Number20140184
Product Code POE
UDI-Device Identifier00380652405850
UDI-Public00380652405850
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P190018
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/20/2026
Device Model NumberCCWET0
Device Catalogue NumberCCWET0.240
Device Lot Number25342107
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/11/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/26/2022
Initial Report FDA Received Date11/23/2022
Date Device Manufactured02/22/2022
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-