| Model Number |
MAJ-1971 |
| Medical Device Problem Code |
Obstruction of Flow (2423)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/31/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Reports are being submitted on the scopes and the clogged connecting tube that were reprocessed using the oer-4.Please refer to the following reports: patient identifier of (b)(6) is related to model number: maj-1971, serial number: unknown patient identifier of (b)(6) is related to model number: jf-260v, serial number: (b)(4).Patient identifier of (b)(6) is related to model number: gif-q260j, serial number: (b)(4).Patient identifier of (b)(6) is related to model number: pcf-q260ji, serial number: (b)(4).The oer-4 was used as the automatic endoscope reprocessor (aer) at the site.This event is under investigation.A supplemental report will be submitted upon receiving additional information.
|
| |
|
Event or Problem Description
|
|
The customer reported the oer-4 automatic endoscope reprocessor (aer) and connecting tube were used to wash a scope.When washing the scope, the connecting tube was clogged and it was possible that the auxiliary water pipe and channel could not be washed.It was unknown if the insufficiently cleaned scope was used on a patient.There was no reported patient harm or impact due to this event.The clogging was resolved and the scope was cleaned again without problems.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's investigation and correction to g2.The device history record was unable to be reviewed for this device since the serial number was not provided due to it being an accessory device.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Based on the results of the investigation, it's likely foreign material entered the connecting tube which led to clogging of the connecting tube.A final root cause of the tube clogging could not be determined.The following is included in the instructions for use: "chapter 8 preparation and inspection¿ of the instruction manual of maj-1971." olympus will continue to monitor field performance for this device.
|
| |
|
Search Alerts/Recalls
|