MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS INC. ATELLICA IM FOLATE (FOL); FOLATE ASSAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS INC. ATELLICA IM FOLATE (FOL); FOLATE ASSAY Back to Search Results
Model Number N/A
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/11/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The outside the us customer reports observation of an elevated patient sample result when using atellica im 1300 folate (fol) assay.After repeating with dilution, the results were significantly lower than the initial result.It is unknown which result is correct.Quality control (qc) levels were back in range after calibration.The interpretation of results section of the atellica im folate (fol) instructions for use states: "results of this assay should always be interpreted in conjunction with patient's medical history, clinical presentation and other findings." siemens healthcare diagnostics is investigating.
 
Event or Problem Description
The customer reports observation of an elevated patient sample result when using atellica im 1300 folate (fol) assay.After repeating with dilution, the results were significantly lower than the initial result.It is unknown which result is correct.There are no known reports of patient intervention or adverse health consequences due to the discordant folate results.
 
Additional Manufacturer Narrative
Initial mdr 1219913-2022-00412 was filed on nov 23, 2022 reporting an outside the us customer reports observation of an elevated patient sample result when using atellica im 1300 folate (fol) assay.After repeating with dilution, the results were significantly lower than the initial result.Additional information - dec 07, 2022.Section d8 - was this device serviced by a third party? updated to no.The customer confirmed the sample type: serum.The initial result was not reported to the physician(s).The repeated (diluted) results were considered as correct by the physician(s).Siemens continues to investigate.
 
Additional Manufacturer Narrative
Siemens filed the initial mdr 1219913-2022-00412 on nov 23, 2022 siemens filed the mdr 1219913-2022-00412 supplemental 1 on dec 19, 2022.Additional information on jan 27, 2023: an outside the united states customer reported observation of elevated patient sample result higher than the measuring interval (>54,36 nmol/l) and after dilution, the result is in range and significantly lower than the initial result.The customer used the reagent lot 339.The sample was a serum sample: the customer uses sarstedt tube, specifically serum gel.The sample was tested directly from the primary tube as well as for repeat/dilution testing.The issue was reproducible with repeat testing with multiple calibrations, multiple ancillary dtt preparations.Siemens determined that the instrument is diluting sample appropriately based on 27 other patient samples diluting as expected.The lab did not determine if the patient was on any medications that could interfere with folserum.There is no evidence of sample integrity errors.The sample had h/i/l calculated to be 0.Siemens concludes that pre-analytical variables can affect the quality of the sample, and deviation from recommended best practices can impact results.While there is insufficient information to determine the cause of the discrepant result, siemens cannot rule out pre-analytical variables or sample specific issues as contributing factors.No product issue has been identified.The customer is operational.In section h6, the investigation finding, and investigation conclusion codes were updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATELLICA IM FOLATE (FOL)
Common Device Name
FOLATE ASSAY
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC.
511 benedict avenue
tarrytown NY 10591 5097
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS INC
333 coney street
east walpole MA 02032
Manufacturer Contact
louise mclaughlin
333 coney street
east walpole, MA 02032
7818564812
MDR Report Key15852262
Report Number1219913-2022-00412
Device Sequence Number15032263
Product Code CGN
UDI-Device Identifier00630414598963
UDI-Public00630414598963
Combination Product (Y/N)N
Initial Reporter CountryEZ
PMA/510(K) Number
K172201
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/29/2023
Device Model NumberN/A
Device Catalogue Number10995572
Device Lot Number339
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 11/14/2022
Supplement Date Received by Manufacturer12/07/2022
01/27/2023
Initial Report FDA Received Date11/23/2022
Supplement Report FDA Received Date12/19/2022
02/21/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/29/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-