| Model Number |
INGENIA 1.5T |
| Medical Device Problem Codes |
Improper or Incorrect Procedure or Method (2017); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
Partial thickness (Second Degree) Burn (2694)
|
| Date of Event |
11/16/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Philips received a repot on a heating incident.Patient claims burn from anterior coil during an mr examination.
|
| |
|
Additional Manufacturer Narrative
|
|
Philips has investigated the reported problem.Philips has completed a good faith effort to get further information regarding the sequence of events of the reported problem.There is no indication of a malfunction of the mr system or coil used that could have contributed to the incident.The observed blistering can be explained by the presence of the ecg electrodes that were left in place after the ct examination that was done before mr examination.The location of the burn marks corresponds with the location of the ecg electrodes.Philips is closing this complaint and this event will be included in the post market surveillance trend analysis.
|
| |
|