| Model Number |
N/A |
| Medical Device Problem Codes |
Separation Failure (2547); Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/29/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4) visual examination of the returned product identified there is a scratch on the inside diameter.When looking at the center hole there is visible damage to the threads, so no further analysis was completed.The handle was not returned, nor pictures provided.Review of the device history record(s) identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2022-02669.
|
| |
|
Event or Problem Description
|
|
It was reported that during a hip procedure, the handle got stuck on the g7 cup.There was no harm or injury to the patient.Attempts have been made and additional information on the reported event is unavailable at this time.
|
| |
|