| Model Number |
LD202 |
| Medical Device Problem Code |
Smoking (1585)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/25/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A steris service technician arrived onsite to inspect the lighting system and found the orbis lighthead board assembly to be burned and was emitting the smell of smoke.The technician replaced the lighthead board assembly, tested the lighting system, confirmed it to be operating according to specifications, and returned it to service.A 3-year complaint review indicates this to be an isolated event.No additional issues have been reported.
|
| |
|
Event or Problem Description
|
|
The user facility reported that prior to the start of a patient procedure smoke was observed emitting from their amsco ld202 in-ceiling lighting system.The patient was transferred to a different room and the procedure was completed successfully.No report of injury or procedure delay.
|
| |
|