• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH PRIMARY TRITANIUM HEMI CLUSTER HOLE CUP 52MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH PRIMARY TRITANIUM HEMI CLUSTER HOLE CUP 52MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Model Number 502-03-52D
Medical Device Problem Code Loss of Osseointegration (2408)
Health Effect - Clinical Code Inadequate Osseointegration (2646)
Date of Event 11/02/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Device not available.
 
Event or Problem Description
Revision report as below: "primary hip done 2 years ago.Patient has complained of pain ever since having it done.Bone scans were conducted.First scan was thought to be fine.Requested a second scan and decided it showed subtle signs of loosening.The cup was removed during the revision ¿very easily¿ and that no bone growth was evident on the cup thus was microscopically loose.".
 
Additional Manufacturer Narrative
Corrected data g1 manufacturing site for devices to sho.An event regarding loosening involving a trident shell was reported.The event was confirmed via clinician review of the provided medical records.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Clinician review: a review of the provided medical information by a clinical consultant indicated: "the revision operative report describes the acetabulum as being clinically loose with no bony ingrowth.Revision tha surgery for loss of/lack of bony ingrowth of the acetabulum is confirmed.The root cause of lack of ingrowth cannot be determined from this limited amount of documentation." product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient was revised due to loosening of the shell.The event was confirmed via clinician review of the provided medical records.The exact cause of the event could not be determined because insufficient information was provided.Further information such as return of the device, pathology reports, pre- and post-operative x-rays as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
Revision report as below: "primary hip done 2 years ago.Patient has complained of pain ever since having it done.Bone scans were conducted.First scan was thought to be fine.Requested a second scan and decided it showed subtle signs of loosening.The cup was removed during the revision ¿very easily¿ and that no bone growth was evident on the cup thus was microscopically loose".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRIMARY TRITANIUM HEMI CLUSTER HOLE CUP 52MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
joann ripoli
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key15852872
Report Number0002249697-2022-01724
Device Sequence Number6525286
Product Code LPH
UDI-Device Identifier07613327040128
UDI-Public07613327040128
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K143085
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/26/2024
Device Model Number502-03-52D
Device Catalogue Number502-03-52D
Device Lot NumberLL8V0L
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 11/02/2022
Supplement Date Received by Manufacturer02/11/2023
Initial Report FDA Received Date11/23/2022
Supplement Report FDA Received Date03/09/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/29/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age71 YR
Patient SexMale
Patient Weight79 KG
-
-