INTRINSIC THERAPEUTICS, INC BARRICAID ACD Back to Search Results
Model Number BAR-A8-10MM
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/26/2022
Type of Reportable Event Malfunction
Event or Problem Description
Initial implantation attempt missed the defect in the disc surface.After a few impacts with the mallet, the nitinol guide wire that holds the implant broke.The implant and wire were removed without any patient harm.A second implantation attempt was successful with a new device.
 
Additional Manufacturer Narrative
A company representative who was present for the implantation commented that the implant did not align with the defect in the disc wall.Additionally, the defect measurement tool was used after the first implantation and revealed that the path was not clear enough for the guide and mesh to be implanted correctly.This combination of factors caused additional force on the guide wire which subsequently broke.There is no risk to the patient according to our hha.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARRICAID ACD
Common Device Name
BARRICAID
Manufacturer (Section D)
INTRINSIC THERAPEUTICS, INC
30 commerce way
woburn MA 01801
Manufacturer (Section G)
INTRINSIC THERAPEUTICS, INC
30 commerce way
woburn MA 01801
Manufacturer Contact
rachael nelsen
30 commerce way
woburn, MA 01801
7819320222
MDR Report Key15854350
Report Number3006232063-2022-00016
Device Sequence Number17721632
Product Code QES
UDI-Device IdentifierM906BARA810MM0
UDI-PublicM906BARA810MM0
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P160050
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/21/2024
Device Model NumberBAR-A8-10MM
Device Catalogue Number2731015-A8
Device Lot Number04302120
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/26/2022
Initial Report FDA Received Date11/23/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/06/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age71 YR
Patient SexFemale
-
-