MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION PLATINUM PLUS; WIRE, GUIDE, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION PLATINUM PLUS; WIRE, GUIDE, CATHETER Back to Search Results
Model Number 39895
Medical Device Problem Codes Entrapment of Device (1212); Stretched (1601); Unraveled Material (1664); Difficult to Advance (2920)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/31/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Initial reporter address 1: (b)(6).
 
Event or Problem Description
It was reported that guide wire entrapment occurred, which prolonged the procedure.The target lesion was located in the inferior branch of left pulmonary artery (lpa).After a 7fr guide catheter was engaged during a cardio mems procedure, a 018/260 platinum plus guidewire was inserted; however, it could not advance.The coil wire unwound from the core and the wire would not traverse the catheter in the right ventricular (tip was in the lpa).The wire core came back easily but the platinum core was still visible.The wire never exited the non-boston scientific balloon tipped catheter and all components were removed from the body.The platinum wire came out in a very long strand and the tip was found to be unraveled and frayed.The procedure was prolonged and was completed with a v-18 guidewire.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PLATINUM PLUS
Common Device Name
WIRE, GUIDE, CATHETER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15854423
Report Number2124215-2022-47161
Device Sequence Number6450532
Product Code DQX
UDI-Device Identifier08714729035961
UDI-Public08714729035961
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K935997
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 11/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/21/2025
Device Model Number39895
Device Catalogue Number39895
Device Lot Number0029633686
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/01/2022
Initial Report FDA Received Date11/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/22/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-