| Model Number |
03.505.006 |
| Medical Device Problem Code |
Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.(b)(4).Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Initial reporter occupation: initial reporter is a synthes employee.Device evaluated by mfr: without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2022 the axel and shaft of the screwdriver in question are stuck together.There was no reported patient interaction, event occurred post-op.No further information is available.This report is for a driveshaft for screwdriver 90°.This is report 1 of 1 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.A manufacturing record evaluation cannot be performed due to lot number being unknown, since the device is embedded inside the mating device and was not possible to disassemble.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned driveshaft for screwdriver 90 was observed wit several deep dents at the gear from the tip.The device was returned with all the components still in assembled condition; however, the components could not be disassembled, this condition is related with the customer complaint.The dimensional inspection was not performed for the driveshaft for screwdriver 90 due to not being applicable to the complaint condition.A functional test was performed to asses the device interaction issue was conducted.The inner shaft could not be removed from the screwdriver shaft.The complaint condition was able to be replicated.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The complaint condition was confirmed for the driveshaft for screwdriver 90 as the device components could not disassembled.The root cause for the reported complaint condition is attributed to a design- related/manufacturing issue.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Drawing/specifications reviewed? yes, reviewed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.D10: date of concomitant therapy is november 1, 2022.
|
| |
|
Search Alerts/Recalls
|
|
|