| Model Number |
SRLP-340 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
11/01/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformance's noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side rupture.The device has been explanted and replaced.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side rupture.The device has been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: d9, h3, h6 device evaluation: based on the device analysis grid, the assessments of the complaint are: rupture: observed, one opening on device in the radius side assessed as a surgical impact opening (shell thickness was within specification).Additional observations: stress marks observed.
|
| |
|