• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: US SURGICAL PUERTO RICO TA; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

US SURGICAL PUERTO RICO TA; STAPLE, IMPLANTABLE Back to Search Results
Model Number TA6048L
Medical Device Problem Codes Failure to Align (2522); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Failure to Anastomose (1028); Unspecified Tissue Injury (4559)
Date of Event 10/07/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Concomitant medical products: ta6048s (ta6048s ta 60-4.8 reloadable stapler, lot number: p2a0279).If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during a colon sigmoid resection appendectomy, two reloads were used.One had malformed staples while the other one had its pin misaligned.The surgeon had to over sew the anastomosis.
 
Additional Manufacturer Narrative
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the retaining pin did not seat properly upon the first squeeze of the handle.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TA
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer (Section G)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key15855093
Report Number2647580-2022-04296
Device Sequence Number5054574
Product Code GDW
UDI-Device Identifier10884521043220
UDI-Public10884521043220
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
K111825
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 02/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2023
Device Model NumberTA6048L
Device Catalogue NumberTA6048L
Device Lot NumberP8F1192MX
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/02/2022
Supplement Date Received by Manufacturer01/27/2023
Initial Report FDA Received Date11/23/2022
Supplement Report FDA Received Date02/10/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/07/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age71 YR
Patient SexFemale
Patient Weight87 KG
-
-