MEDTRONIC HEART VALVES DIVISION EVOLUT PRO TRANSCATHETER AORTIC VALVE; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV
|
Back to Search Results |
|
Model Number EVOLUTPRO-29-US |
Device Problem
Material Split, Cut or Torn (4008)
|
Patient Problems
Cusp Tear (2656); Aortic Valve Insufficiency/ Regurgitation (4450)
|
Event Date 10/28/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
Product analysis: no product was returned.Conclusion: without the return of the product, no definitive conclusion can be made regarding the clinical observation.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information that approximately three years and eleven months following the implant of this transcatheter bioprosthetic valve, the patient was evaluated for a transcatheter valve replacement due to unspecified aortic insufficiency from a leaflet tear.A 26 mm transcatheter valve was implanted valve in valve.No additional adverse patient effects were reported.
|
|
Event Description
|
Additional information was received that the cause of the leaflet tear was unknown.It was noted a post-implant balloon aortic valvuloplasty was not performed following valve implant.The severity of the aortic insufficiency was severe.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Conclusion: the device history record was reviewed and showed that this product met all manufacturing specifications for product released for distribution.No issues were identified that would have impacted this event.Leaflet tear is a known potential adverse effect per the device instructions for use (ifu).It was reported that the cause of the leaflet tear was unknown but a post-implant balloon aortic valvuloplasty was not performed after valve implant.Thus, with the limited information available, a conclusive root cause could not be determined.Regurgitation can be caused by a variety of factors, including valve positioning, patient anatomy, or presence of pre-existing patient conditions.With the limited information available, a conclusive root cause could not be determined but the leaflet tear was a likely contributing cause.There is no information to suggest a device quality deficiency that may have caused or contributed to this event.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|