| Model Number |
1312-18-303 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Pain (1994); Foreign Body In Patient (2687)
|
| Date of Event |
10/26/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Foreign source: france.Customer has indicated that the product will not be returned to zimmer biomet for investigation, as the device remains implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent a placement of a posterolateral distal humerus plate.The surgeon noticed during the follow up x-ray that the fast guides were left inside the patient.A revision surgery was performed the day after implantation to remove the fast guide.No additional patient consequences were reported.
|
| |
|
Event or Problem Description
|
|
No additional event information is available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated: b4, b5, d1, d2, d4, g3, g4, g6, h1, h2, h10 the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.Device history record (dhr) review was unable to be performed as the lot number of the device involved in the event is unknown.The root cause of the reported issue is attributed to use error as the surgical technique indicates to remove f.A.S.T.Guide inserts after drilling.A summary of the investigation has been sent to the complainant.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.H3 other text : device remains implanted.
|
| |
|
Event or Problem Description
|
|
No additional event information is available.
|
| |
|