| Model Number |
TRLP-300 |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
08/01/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.The event of "capsular contracture, baker grade iv" is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture, baker grade iv.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "bilateral exchange from textured to smooth due to the patients concern of the product and left side capsular contracture, baker grade iv".Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the device analysis grid, the assessments of the complaint are: capsular contracture: unable to observe as it is a medical event not related to the device.Anxiety-product/procedure: unable to observe as it is a medical event not related to the device.Additional observations: one opening observed on posterior side assessed as unidentified (tear) opening (shell thickness was within specification).Brown particles observed on the surface of the shell.Creases were observed.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "exchange from textured to smooth due to the patients concern of the product and left side capsular contracture, baker grade iv".Device has been explanted.
|
| |
|