The device history record for the manufacturing lot was reviewed and there were no discrepancies or unusual findings.A 61-year-old male patient underwent uncomplicated, cataract surgery with well-positioned, in-capsule implantation of a +23.0 diopter lal (sn: (b)(4) in the right eye on (b)(6) 2021.Examination on postoperative day 1 (pod1) showed findings consistent with tass (toxic anterior segment syndrome), which was also reported in 3 other eyes operated the same day using iols from different manufacturers.The tass outbreak was potentially linked to a pharmaceutical lot common to all cases.Despite aggressive management of tass, the eye required descemet stripping automated endothelial keratoplasty (dsaek) two months after surgery.The patient admitted to non-compliance with uv protection following dsaek procedure leading to a central zone of polymerization from ambient light exposure.The lal was explanted and exchanged with a different iol six months after surgery.The explanted lal was not returned to rxsight, but was analyzed by the university of utah ophthalmic pathology group, which noted a paracentral elevation on the lal.Manufacturer reference #: (b)(4).
|