FISHER & PAYKEL HEALTHCARE LTD OPTIFLOW JUNIOR INTERFACE; BTT Back to Search Results
Model Number OPT318
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/08/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The optiflow junior nasal cannula is designed specifically for the delicate anatomical features and flow requirements of neonatal and paediatric patients.It features adhesive pads (wigglepads) to maintain cannula stability on the patient's cheeks, soft-touch nasal prongs and breathable kink-proof and crush-resistant flexible tubing.Method: the complaint opt318 optiflow junior nasal cannula was not returned to fisher & paykel healthcare (f&p) for evaluation.Our investigation is thus based on the information and photograph provided by the customer and our knowledge of the product.Result: visual inspection of the provided photograph revealed that the cannula was broken between the nasal prongs.It was also observed that one of the prongs was missing.Conclusion: we are unable to determine the cause of the reported event.All optiflow junior cannula are 100% leak and occlusion tested after final assembly and any cannula that fails is discarded.In addition, samples are taken hourly from each run and pull tested to check glue joint strength at the cannula/tube joint, as well as the swivel grip joint.If there are any failures the entire batch is put aside for further investigation.The subject opt318 optiflow junior nasal cannula would have met the required specification at the time of production.The user instructions which accompany the opt318 optiflow junior nasal cannula show in pictorial format the correct placement and fitting of the cannula and also warn: ensure that all connections are secure during use.Check cannula is undamaged and that the flow path is maintained.Under excessive load, the cannula may disconnect to prevent forces being transferred to the patient.Appropriate patient monitoring must be used at all times.Failure to monitor the patient may result in loss of therapy, serious injury or death.Do not stretch or crush tube.
 
Event or Problem Description
A distributor in china reported behalf of a healthcare facility that the opt318 optiflow junior nasal cannula was broken between the nasal prongs.There was no reported patient consequence.
 
Event or Problem Description
A distributor in china reported behalf of a healthcare facility that the opt318 optiflow junior nasal cannula was broken between the nasal prongs.There was no reported patient consequence.
 
Additional Manufacturer Narrative
(b)(4).Corrections: the following corrections have been made.D9: the device was returned to manufacturer for further investigation.H3: device is evaluated by the manufacturer.H6: adverse event problem; type of investigation.The optiflow junior nasal cannula is designed specifically for the delicate anatomical features and flow requirements of neonatal and paediatric patients.It features adhesive pads (wigglepads) to maintain cannula stability on the patient's cheeks, soft-touch nasal prongs and breathable kink-proof and crush-resistant flexible tubing.Method: the complaint opt318 optiflow junior nasal cannula was returned to fisher & paykel healthcare (f&p) for evaluation.The affected cannula was visually inspected and an analysis was done through scanning electronic microscope (sem).Result: visual inspection of the returned device revealed that the cannula was broken between the nasal prongs.Also, it was revealed that the right nasal prong was missing.The sem analysis revealaed that there was extensive evidence of tearing.Conclusion: we are unable to determine the cause of the reported event.All optiflow junior cannula are 100% leak and occlusion tested after final assembly and any cannula that fails is discarded.In addition, samples are taken hourly from each run and pull tested to check glue joint strength at the cannula/tube joint, as well as the swivel grip joint.If there are any failures the entire batch is put aside for further investigation.The subject opt318 optiflow junior nasal cannula would have met the required specification at the time of production.The user instructions which accompany the opt318 optiflow junior nasal cannula show in pictorial format the correct placement and fitting of the cannula and also warn: - ensure that all connections are secure during use.Check cannula is undamaged and that the flow path is maintained.Under excessive load, the cannula may disconnect to prevent forces being transferred to the patient.- appropriate patient monitoring must be used at all times.Failure to monitor the patient may result in loss of therapy, serious injury or death.- do not stretch or crush tube.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTIFLOW JUNIOR INTERFACE
Common Device Name
BTT
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
faranak gomarooni
17400 laguna canyon road
suite 300
irvine, CA 92618
9494534002
MDR Report Key15859525
Report Number9611451-2022-01099
Device Sequence Number12223789
Product Code BTT
UDI-Device Identifier09420012432308
UDI-Public(01)09420012432308(10)2101789577(11)210906
Combination Product (Y/N)N
Initial Reporter CountryCH
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Distributor
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 11/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberOPT318
Device Catalogue NumberOPT318
Device Lot Number2101789577
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/11/2022
Supplement Date Received by Manufacturer03/19/2023
Initial Report FDA Received Date11/23/2022
Supplement Report FDA Received Date03/27/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/06/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-