| Model Number |
UNKNOWN |
| Medical Device Problem Code |
Structural Problem (2506)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/28/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Initial reporter phone #:(b)(6).A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that while using the bd kiestra¿ inoqula+¿ tla that the barcoda hood was falling.The following information was provided by the initial reporter: replace barcoda hood gas springs.
|
| |
|
Additional Manufacturer Narrative
|
|
H.6 investigation summary: bd technical services reported a worn-out gas spring on instrument bd kiestra inoqula tla (material:447213, serial number: (b)(6)).Springs were replaced with new set which has confirmed the complaint.Root cause is defined as faulty springs, and replacement with new set is initiated as a corrective action.Design history record (dhr) review is not required for this complaint, the complaint was evaluated via other elements of the investigation.The results of this evaluation have not identified any new hazards, new risks, or specific trends.
|
| |
|
Event or Problem Description
|
|
It was reported that while using the bd kiestra¿ inoqula+¿ tla that the barcoda hood was falling.The following information was provided by the initial reporter: replace barcoda hood gas springs.
|
| |
|