We checked the returned unit and confirmed that the ccd module blackout.Based on the result, we concluded that it was caused due to the physical damage applied on the ccd module.In addition, we confirmed that the angle wire fluid damage, the insertion flexible tube (ift) fluid damage, the control body fluid damage, the down pulley wire fluid damage, the ccd module fluid damage, the electrical connector fluid damage, the lg cable connector fluid damage, the shield cover fluid damage, the control body corroded, the receptacle corroded, the lg cable connector corroded, the shield cover corroded, the operation channel leak, the distal body worn out, and the insertion flexible tube (ift) bump; however, they are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report "hr-rpt-0586(image failure)" and/or the risk analysis results, it was evaluated to submit mdr.
|