MAUDE Adverse Event Report: BECTON, DICKINSON & CO. (SPARKS) BD BBL¿ MACCONKEY II AGAR; CULTURE MEDIA, SELECTIVE AND DIFFERENTIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON & CO. (SPARKS) BD BBL¿ MACCONKEY II AGAR; CULTURE MEDIA, SELECTIVE AND DIFFERENTIAL Back to Search Results
Model Number 221270
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/09/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.Initial reporter address 1: (b)(6).
 
Event or Problem Description
It was reported that bd bbl¿ macconkey ii agar contamination occurred.The following information was provided by the initial reporter: 30 packs of contaminated plated media.
 
Event or Problem Description
It was reported that bd bbl¿ macconkey ii agar contamination occurred.The following information was provided by the initial reporter: 30 packs of contaminated plated media.
 
Additional Manufacturer Narrative
H.6 investigation summary: during manufacturing of material 221270, media is formulated and sent through a high temperature short time sterilizer to remove bioburden.The petri dishes are subjected to uv radiation to decrease bioburden.The petri dishes are filled in a positive pressure hepa filtered environment.The filled plates are cooled and immediately wrapped into sleeves to decrease the introduction of microbes.Sleeves are then packaged into cartons and then transferred to a refrigerated truck (2 to 8 degrees c) for shipment to the distributor.Bd distributors are provided with the storage guidelines for the shipping and handling of bd media of 2 to 8 degrees c in a dark place.The batch history record review for batch 2214274 was satisfactory per internal procedures.The release testing that is performed on this product does include bioburden testing.A sample of plates are incubated at 25 degrees c and at 35 degrees c for approximately 72 hours.All bioburden testing performed on this batch was satisfactory per bd internal procedures.Affected product does not have any sterility claims; the product is tested for bioburden prior to release to ensure that it conforms to product specifications.However, this does not ensure that the end-user will not receive a contaminated plate.The complaint history was reviewed, an no other complaints have been taken on batch 2214274.No retention samples for this batch were available for investigation.Two photos were received.Photo one shows sleeve lot 2215274, no defect seen in the photo.Photo two shows open sleeve of plates.Top plate (id 2214274 timestamp not visible) shows multiple colonies.No return samples were received for investigation of this complaint.This complaint can be confirmed.Bd has identified a contamination trend for this product and the investigation found opportunities for bioburden reduction in the manufacturing process.Additional trainings are planned with an ongoing training review for cleaning processes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD BBL¿ MACCONKEY II AGAR
Common Device Name
CULTURE MEDIA, SELECTIVE AND DIFFERENTIAL
Manufacturer (Section D)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer (Section G)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15863862
Report Number1119779-2022-01415
Device Sequence Number20329081
Product Code JSI
UDI-Device Identifier10382902212703
UDI-Public10382902212703
Combination Product (Y/N)N
Initial Reporter CountryRP
PMA/510(K) Number
NA
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date10/31/2022
Device Model Number221270
Device Catalogue Number221270
Device Lot Number2214274
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/09/2022
Supplement Date Received by Manufacturer02/14/2023
Initial Report FDA Received Date11/24/2022
Supplement Report FDA Received Date02/20/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/02/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-