• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER GMBH 6.5MM CENTER SCREW - 36M M; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH 6.5MM CENTER SCREW - 36M M; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED Back to Search Results
Model Number 5573-6536
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Bone Fracture(s) (1870); Implant Pain (4561)
Date of Event 11/24/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Based on the available information the device will not be returned therefore an evaluation of the device cannot be performed.A review of the device history is not possible because the lot number was not communicated.Should additional information become available, it will be provided in a supplemental report.
 
Event or Problem Description
As reported: "acromial stress fracture non-union.Patient experiencing pain over proximal and distal locking bolt sites.Reoperation to remove locking bolts".
 
Event or Problem Description
As reported: "acromial stress fracture non-union.Patient experiencing pain over proximal and distal locking bolt sites.Reoperation to remove locking bolts".
 
Additional Manufacturer Narrative
The reported event could be confirmed based on the x-rays received for investigation.The inspection of the pictures revealed the following: the post-operative x-rays confirmed the reported pain due to acromial stress fracture occurring 6 months after implantation.However, it was not possible to determine the specific cause of the post-operative fracture.To determine a root cause in this case, more clinical information is needed.Medical experts have indicated that: " the incidence of acromial stress fractures in reversed shoulder arthroplasty is relatively rare but has been reported in a small number of case studies and case series.Preventative measures include careful patient selection, appropriate implant positioning, and postoperative rehabilitation to prevent overuse of the affected shoulder.Acromial stress fractures in reversed shoulder arthroplasty can be caused by several factors, including: 1.High levels of stress on the acromion due to abnormal biomechanics caused by the reversed shoulder implant.2.Pre-existing osteoporosis or other bone conditions that make the acromion more susceptible to fractures.3.High activity levels or repetitive overhead motions that increase stress on the acromion.4.Postoperative infection or inflammation that can lead to increased stress on the acromion.5.Surgical technique and implant design, that may cause increased stress on the acromion.It is important to note that these are not the only possible causes of acromial stress fractures, and further evaluation by a medical professional is necessary to determine the specific cause of the fracture." based on the investigation, it was not possible to determine the root cause of the reported event.Several factors may lead to an acromial stress fracture and in order to determine the specific cause in this case, more clinical information is needed.A review of the device history was not possible because the lot number was not communicated.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If the device is returned or if any additional information is provided, the investigation will be reassessed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
6.5MM CENTER SCREW - 36M M
Common Device Name
PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key15864177
Report Number0008031020-2022-00645
Device Sequence Number14950861
Product Code KWS
UDI-Device Identifier07613327098846
UDI-Public07613327098846
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K202289
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 03/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number5573-6536
Device Catalogue Number5573-6536
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/31/2022
Supplement Date Received by Manufacturer02/20/2023
Initial Report FDA Received Date11/25/2022
Supplement Report FDA Received Date03/18/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age72 YR
Patient SexMale
Patient Weight84 KG
-
-