|
H10: the medical device manufacturer (d3) and manufacturing location (g1) for the straub product was selected as unknown due to system limitations.The correct medical device manufacturer and manufacturing location are straub medical us.H10: manufacturing review: a manufacturing review was conducted and there was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the physical sample was returned for evaluation.A physical investigation was performed for the catheter.Investigated broken helix and tube kinked but not fully detached at 20 cm from the tip of the catheter.Therefore, it can be confirmed that the helix was broken.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|