Model Number P8040H000495
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/21/2022
Type of Reportable Event Malfunction
Event or Problem Description
Hillrom received a report from a hillrom technician stating the bed brakes were not holding.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint #(b)(4).
 
Additional Manufacturer Narrative
The hillrom technician found the casters needed to be replaced.Per the¿hillrom¿service manual the bed should be subject to an effective maintenance program.¿ an annual service of the bed is advised in order to maintain its characteristics and performance.¿ brake casters should be checked for cuts, wear and quality of tread, etc.And replaced when necessary.¿ apply the brake and make sure the stretcher does not move.If there is movement, look at the brake components for wear.¿ apply the steer, and make sure the stretcher steers correctly.¿ look at the steer components for wear.¿ put the stretcher in neutral.Make sure all four casters rotate and roll freely.¿¿adjust or replace components if necessary.¿ a search of the hillrom maintenance records did not show hillrom performed any preventative maintenance on this bed.¿ it is unknown if the facility performs preventative maintenance on their beds.¿¿ the technician replaced the casters to resolve the reported event.¿¿based on this information, no further action is required.¿.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRAUMA STRETCHER FRAME
Common Device Name
STRETCHER, WHEELED
Manufacturer (Section D)
HILL-ROM MEXICO
ave.del telefono no. 200 col. huinala
apodaca, nuevo leon 66640
MX  66640
Manufacturer Contact
maritza valencia
1069 state route 46 east
batesville, IN 47006
8129310130
MDR Report Key15865027
Report Number3006697241-2022-00084
Device Sequence Number6995978
Product Code FPO
UDI-Device Identifier00887761000995
UDI-Public010088776100099511190911
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K961437
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 11/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberP8040H000495
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/21/2022
Initial Report FDA Received Date11/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/10/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-