| Model Number |
SRL-230 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
11/05/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Health care professional reported ride side rupture.Device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: b5, d6b, d9, h3, h6.
|
| |
|
Event or Problem Description
|
|
Device has been explanted.
|
| |
|
Event or Problem Description
|
|
Health care professional reported ride side rupture.Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the device analysis grid, the assessments of the complaint are: rupture: observed, opening on the posterior side assessed as surgical impact opening.(shell thickness was within specification).Additional observations: no penetrating nick ( stress marks) observed on the device.No further actions are required as the device was implanted.
|
| |
|