| Catalog Number |
05342716001 |
| Medical Device Problem Codes |
Circuit Failure (1089); Sparking (2595)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/07/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Further investigation is needed to determine the root cause(s) of the alleged issue.No harm alleged.The pcba cell bus was replaced and the instrument is working as intended.
|
| |
|
Event or Problem Description
|
|
A field specialist observed evidence of charring/burning of the cell bus printed circuit board (pcb) on the benchmark ultra instrument for a customer from the netherlands.No harm or injury alleged.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation concluded that the service action resolved the issue.Updated and replaced the device, evaluation method and evaluation result and conclusion codes.
|
| |
|