H10: as this malfunction is considered one event identified by the dealer; only one mdr report will be submitted for the reported quantity affected of twenty for this event.H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one bard ptfe pledget package was returned for evaluation.During visual evaluation, no evidence of breach of sterile barrier was observed.Sample was noted to have a black particle within packaging along with discoloration.Therefore, the investigation is confirmed for the reported contamination issue.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 08/2026), g3, h6 (method).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|