Product analysis the device was returned with the manifold safety locking pin loose.The device was returned in a deployed state with the grips pushed together the device was returned with the distal end of the outer sheath detached from the device, the nylon inner shaft was broken and the inner detached approximately 5mm proximal to retainer, there was several kinks / bends on the detached piece of outer sheath the device returned with a kink on the blue outer sheath, approx.16.2 cm distally from the strain relief no functional testing could be carried out due to the condition of the returned device.The distal inner assembly was cut just proximal to the stent retainer to allow further testing.A set of witness marks were noted on the stainless steel hypotube approximately 27mm and 29mm from the distal end of the stainless steel hypotube; indicating that the safety locking pin was properly tightened during manufacturing.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|