The user facility reported that the tip of the involved navicross broke off in the patient.Doctor stated that the patient was ok.The event occurred inta-operative.The procedure outcome was completed.The patient was not injured, and surgical intervention was not required.Additional information was received on (b)(6) 2022 : procedure taking place, when the tip of the navicross broke off inside the patient was pad below the knee case.The navicross broke off below the knee and were able to remove the navicross tip from the patient's anatomy.Surgical intervention was not required.There was no tortuous anatomy experienced during the procedure.The original procedure was completed as intended.
|
|
Udi: lot number was unknown, di no: (b)(4).Implanted date: device was not implanted.Explanted date: device was not explanted.Phone number: requested, unknown.The actual device was not returned; therefore, an evaluation of the actual device was unable to be conducted.Since the lot number was unknown, the manufacturing record and the shipping inspection record could not be reviewed.Information on products with the involved product code in the past three (3) years (november 2019 - as of november 2022) were investigated.As a result, it was confirmed that no deviations or nonconformities related to the involved event had occurred.A search of the complaint file in the three (3) years (november 2019 - as of november 2022) found no other similar report with the involved product code from other facilities.Based on the investigation result, for products of the involved product code in the past, no deviations or nonconformities related to the involved event had occurred, and no other similar report from other facilities were found.Since the actual sample and cine image were not sent, the cause of occurrence could not be clarified.The instructions for use (ifu) has the following warning and directions for use.Warnings and precaution - carefully handle the product under fluoroscopy.If any resistance is felt while handling the product, immediately stop the manipulation and find out the cause to avoid damage to blood vessels and separation or breakage of the product.Directions for use - 3.Advance the guide wire and the product to the target vessel under fluoroscopy.Please see mdr 2243441-2022-00030 for the importer report.Terumo medical products (tmp) (importer) registration no.2243441 is submitting this report on behalf of ashitaka factory of terumo corporation (manufacturer) registration no.9681834.
|