• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER TRAUMA KIEL DRILL, TRI-FLAT T2 TIBIA Ø3,5X130 MM; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER TRAUMA KIEL DRILL, TRI-FLAT T2 TIBIA Ø3,5X130 MM; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT Back to Search Results
Model Number 1806-3555
Medical Device Problem Codes Break (1069); Fracture (1260)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 11/02/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Once the investigation has been completed any additional information will be reported in a supplemental report.
 
Event or Problem Description
As reported: "a drill breakage occurred during freehand drilling for insertion of a distal screw.The broken drill tip was deemed impossible to extract and was left in the patient.".
 
Additional Manufacturer Narrative
The reported event could be confirmed, since the device was returned, and matches the alleged failure.The device inspection revealed that: the visual inspection of the drill, tri-flat t2 tibia ø3,5x130 mm confirmed that the tip of the drill broke off.The distal end of the drill shows abrasive marks which is evident from excessive usage over time.The microscopic images show a brittle fracture which is evident from the shiny surface, which is the result of excessive torsional forces.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.Based on the investigation, the root cause was attributed to a user related issue.The failure was caused by the application of excessive torsional force applied by the user.If the device is returned or if any additional information is provided, the investigation will be reassessed.
 
Event or Problem Description
As reported: "a drill breakage occurred during freehand drilling for insertion of a distal screw.The broken drill tip was deemed impossible to extract and was left in the patient.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRILL, TRI-FLAT T2 TIBIA Ø3,5X130 MM
Common Device Name
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
Manufacturer (Section D)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM  D-24232
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key15867998
Report Number0009610622-2022-00555
Device Sequence Number5210040
Product Code HWE
UDI-Device Identifier04546540196668
UDI-Public04546540196668
Combination Product (Y/N)N
Initial Reporter CountryJA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1806-3555
Device Catalogue Number18063555
Device Lot NumberKU105860
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/16/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/02/2022
Supplement Date Received by Manufacturer12/22/2022
Initial Report FDA Received Date11/28/2022
Supplement Report FDA Received Date01/18/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/03/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age73 YR
Patient SexMale
-
-