Model Number VAMF4040C200TU
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 10/09/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Loss of distal seal with apparent distal endoleak was observed at the level of the distal stent graft margin of the valiant captivia devices already implanted in the patient prior to the (b)(6) 2022 procedure.It is most likely that aneurysm morphology changes due to disease progression with aneurysm morphology changes over the 6 years of implantation of the devices was a factor in the occurrence of the observed loss of seal leading to a distal endoleak.Analysis of the returned films did not reveal any stent graft integrity issues.Concomitant medical products: other relevant device(s) are: product id: v amc4040c200tu, serial/lot #: (b)(4), ubd: 10-may-2017, udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
A valiant captivia was intended to be implanted during the endovascular treatment of a 67mm thoracic aortic aneurysm.It was reported that during the index procedure, the external iliac artery was torn upon insertion of the device.It was noted that the patient's blood pressure dropped until the pelvis was opened and the bleeding was controlled.The damaged vessel was surgically repaired.Per the physician, the cause of the event was anatomy related to a small external iliac artery.No additional sequelae were reported and the patient is fine.It was reported that the stent graft vamc4238c150tu/sn (b)(4) was not implanted in the patient.Another stent graft was opened and deployed.There were no issues with any other grafts used that day.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALIANT CAPTIVIA - FF
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TR
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
EI 
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
EI  
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key15868028
Report Number9612164-2022-04440
Device Sequence Number17457483
Product Code MIH
UDI-Device Identifier00643169346543
UDI-Public00643169346543
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P100040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2017
Device Model NumberVAMF4040C200TU
Device Catalogue NumberVAMF4040C200TU
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/10/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/04/2022
Initial Report FDA Received Date11/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/01/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age74 YR
Patient SexFemale
-
-